FAQ
A HERD filter is a low-pass filter for the wiring of superconducting and spin qubits. It blocks infrared radiation, which breaks Cooper pairs and limits qubit coherence. The signals that control and read out the qubit pass with ultra-low loss, so readout keeps its speed and fidelity. HERD stands for High-Energy Radiation Drain: the filter is based on a leaky coaxial waveguide, which directs the radiation out of the signal path and into an absorptive region.
HERD filters were invented by the founders of Sweden Quantum and are made by the company. Robert Rehammar and Simone Gasparinetti first published the technology at Chalmers University of Technology in 2022, and Sweden Quantum was spun out of Chalmers to commercialize it. The company holds the patents and continues to develop the filters.
In an Eccosorb filter, the epoxy absorbs the radiation along the signal path, and the signal travels through the same epoxy. HERD directs high-frequency radiation away from the main propagating path of the filter and into an absorptive region. The group of M. Devoret at Yale measured the charge-parity switching rate of a transmon. With Eccosorb CR-110 in a light-tight shield, the rate decreased with filter length, to 0.14 per second at 6.3 cm, at about 13 dB of insertion loss. With well-thermalized HERD in a similar shield it reached 0.2 per second, at about 0.3 dB, and they note that, unlike Eccosorb, HERD has very low insertion loss at 10 GHz.
Absorptive filters introduce in-band loss, and LC low-pass filters leak at high frequencies through parasitic elements. HERD directs infrared radiation out of the coaxial line and into an absorber; signals below the cutoff of the openings stay on the line. HERD has no parasitic leakage paths, so its stopband is not bounded by parasitics the way an LC filter's is. HERD-2 blocks from 70 GHz upward, with >60 dB typical attenuation from 70 to 120 GHz, and keeps insertion loss to 0.25 dB typical from DC to 8 GHz.
Yes. At Cooper-pair-breaking frequencies, neither attenuators nor isolators provide enough IR suppression on the readout. Even if attenuators worked at 100 GHz as well as in band, one estimate still puts the rate through a 60 dB line with no IR filter at about 1,400 pair-breaking photons per second. For scale, the best setups see one parity switch per second or fewer. Other lines have even less protection: flux lines usually have no attenuator at the coldest stage, since bias current would heat it, and the readout output has none at all, to keep its signal.
Install HERD filters below the cryostat base plate, as close as possible to the QPU, and on every line connected to it. HERD has no magnetic parts that could disturb the qubit, so it can sit close to the chip, inside the shield. In setups using TWPA as the first amplifier, we recommend installing a HERD filter as close as possible to the device under test, before the first component in the TWPA solution.
To verify that HERD improves the performance of your quantum computing setup, we suggest the following experiments. (1) Measure quasiparticle tunneling rates across a Josephson junction, for example, by using a charge-sensitive transmon qubit. If the tunneling rates are limited by quasiparticle generation or photon-assisted tunneling due to high-frequency radiation in the control line, you will see an improvement after installing HERD. To observe an improvement, an IR-tight sample enclosure is typically required. (2) Measure coherence times and thermal population of a superconducting qubit with and without HERD installed. The group of M. Devoret (Yale University) has explicitly demonstrated a reduction in the quasiparticle tunneling rate in charge-sensitive transmons from 10 Hz to below 1 Hz using HERD-1 and HERD-2 filters, consistently across multiple experiments.
Yes. HERD-2 is built from non-magnetic materials, so it can be mounted close to both flux-sensitive superconducting qubits and magnetically sensitive spin qubits.
HERD-2 has a gold-plated copper body with mounting holes for mechanical or thermal support. Thermalizing the outer body of HERD-1 with copper braids has been shown to improve performance. Contact us for integration guidance for your cryostat.
We have measured the stopband up to 145 GHz, the limit of our measurement setup. Above that we expect it to continue, since HERD does not rely on resonance and has no parasitic leakage paths, and radiation above 145 GHz is still above the cutoff of the openings.
Insertion loss and attenuation figures are typical values measured at room temperature. The filters are rated from 0 K to 320 K (absolute maximum), and the setups on our research page use them at millikelvin temperatures.
Yes. HERD is precision machined, with low unit-to-unit variation. We test every filter individually before it ships. HERD is in the setups of 30+ papers from 15+ institutions.
Other questions: info@swedenquantum.com